Remote Particle Sensors Models 3718-A and 3719-A

Remote laser particle sensors for continuous fixed-point cleanroom monitoring — 0.3 µm sensitivity, serial (3718-A) or Ethernet (3719-A) network interface for integration with CRMS and BMS.

Size sensitivity
0.3 µm
Channels
4 channels
Flow rate
0.1 CFM (2.83 LPM)
Interface
RS-485 / Ethernet
0.3 µm fixed sensorRS-485 / Ethernet4 channelsContinuous
ISO 14644-2ISO 21501-421 CFR Part 11

Products

ProductProduct lineApplication areaSelection note
Remote Particle Sensors Models 3718-A and 3719-ARemote & Continuous Particle MonitoringParticle Counting & MonitoringSelect options, accessories and calibration scope against the measurement procedure.

Key Features

Particle measurement focus

Supports particle count, particle-size or monitoring tasks in controlled environments.

Portable or fixed deployment

Selection covers handheld checks, portable certification support, remote sensors or continuous monitoring.

Data and alarm workflow

Data export, alarms, communication interface and software depend on selected model.

Cleanroom monitoring fit

Suitable for troubleshooting, routine checks, trend observation and contamination-control monitoring.

Sampling-plan dependent

Flow rate, particle channels and sample location must match the measurement plan.

Calibration planning

Calibration certificate type and service interval should be confirmed before deployment.

Technical Specifications

ParameterUnitValue / Description
PARTICLE MONITORING
Particle channelsTwo-channel remote particle sensor
Flow rate optionsCFM0.1 CFM and 1.0 CFM variants
Particle-size optionsµm0.3 / 0.5 µm or 0.5 / 5.0 µm variants depending on model
IntegrationRS-485 Modbus communication; 4–20 mA analog output on 1.0 CFM unit
OutputsParticle counts, error status and alarm signal
Use modeStandalone, PLC, data logger, process monitoring or building automation integration
CalibrationISO 21501-4 and JIS B9921 per manufacturer material

Values are summarized for page-level selection. Confirm final model configuration, accessories and calibration certificate before order placement.

Applications

Cleanroom particle checks

Supports cleanroom particle checks through appropriate sampling flow, particle channels, data logging and calibration scope.

Certification support

Supports certification support through appropriate sampling flow, particle channels, data logging and calibration scope.

Remote monitoring points

Supports remote monitoring points through appropriate sampling flow, particle channels, data logging and calibration scope.

Troubleshooting and trend checks

Supports troubleshooting and trend checks through appropriate sampling flow, particle channels, data logging and calibration scope.

Semiconductor and electronics cleanrooms

Supports semiconductor and electronics cleanrooms through appropriate sampling flow, particle channels, data logging and calibration scope.

Controlled-area monitoring

Supports controlled-area monitoring through appropriate sampling flow, particle channels, data logging and calibration scope.

Similar Products